Issues with the EEG or Spectral Display

It is most likely the electrodes that are causing these issues.  Sometimes electrodes can develop breaks in them but you will still receive a green impedance reading.  This is due to the break in the electrode connecting while conducting the impedance reading.  If you move the electrodes in a manor that disconnects the break in the electrodes, this can cause irregularities in the EEG and/or Spectral Display.  Some issues you may be experiencing include:  a flat line in your EEG/Spectral Display, erratic lines in the EEG/Spectral Display, or think fuzzy lines in the EEG/Spectral Display.

Please follow the instructions below to troubleshoot your electrodes:

  1. Remove the jumper for the test
  2. Use 5 electrodes
  3. Put the sensor ends of the electrodes into distilled water (add a tiny bit of paste to the water to make it conductive).
  4. Check impedance and observe that you get all green lights.
  5. Run Cygnet in live mode and observe that you get totally flat lines. 
  6. Touch or wiggle each electrode one by one. While the raw trace should deflect a bit from flat, it should not "fuzz-out"
  7. Find the electrode(s) that are causing the signal to behave wildly on the screen when the wires are touched, and set them aside to be returned if they are within 4 months old, or throw them away if they are older than that
  8. Replace with brand new sensors and observe that the signal does not deflect as wildly when the sensors are touched.